Veloce 50 Operating Manual
4
2.3.3 ANALOG OUTPUT
The analog output gives a voltage proportional to the value of the physical parameter measured by the
fiber optic transducer. The analog output is a BNC-type coaxial connector. See section 6 for converting
the voltage output value to the measured physical parameter value.
2.3.4 R
ESET BUTTON
The Reset Button zeroes out the analog output of the Veloce Module. The «Reset» button has to be
pressed each time a transducer is connected to the Transducer Input. When the Reset Button is
pressed, all reference to the previous measurements is lost, thus making the Veloce a relative
measurement instrument. The Veloce module takes approximately 1 second to settle after the Reset
Button has been pressed.